Neuron Spectrum 65 with Polysomnograph

Description

Ready-made solution for your PSG lab

For more than a quarter of a century, Neurosoft has been designing and producing various equipment for electrodiagnostics and neurophysiology. Advanced noise reduction technologies implemented in our systems ensure high-quality of the recorded signal making our products a popular choice among healthcare professionals worldwide.

Today, we offer our Neuron-Spectrum-65/PSG system — a ready-made solution for your PSG lab. The system is intended for long-term synchronous PSG, video and audio data recording as well as for analysis of sleep, cardiorispiratory events and limb movements.

Full-featured PSG recorder of expert class, Neuron-Spectrum-65/PSG allows performing any PSG study according to the latest international requirements. Routine PSG, multiple latency sleep test, maintenance of wakefulness test are possible with that multitask device and simple thanks to the automatic calculation of the parameters necessary for diagnosis.

Full range of PSG channels in compliance with AASM* recommendations

The system is provided with the full range of channels required to perform PSG studies.

Channels for sleep staging

  • up to 32 EEG channels
  • 2 EOG channels
  • 3 EMG channels

Channels for analysis of sleep-related respiratory events

Channels to connect:

  • oronasal thermal airflow sensor
  • nasal pressure airflow sensor (cannula)
  • thorax respiratory effort sensor
  • abdominal respiratory effort sensor
  • snoring sensor
  • body position sensor

Built-in channels:

SpO2 and PPG (photoplethysmography)

  • ECG
  • 8 DC channels for the most demanding PSG studies

Channels for limb movement recording

  • 2 EMG channels

Exellence

Full range as type I sleep monitor with Full Feature PSG software
Full range of PSG channels in compliance with AASM recommendations
Portable patient unit for quick connection/disconnection
Synchronous video monitoring
Manual, semi-automatic and automatic sleep staging and PSG event detection
DC channel for data transfer to Neuron-Spectrum.NET software during CPAP titration
Software EEG & Brain Mapping including
Fitur Utama
PSG Trends:
SpO2, Respiratory, Abdominal Effort, Thorax Effort, Heart Rate, Limb Movement,
Snoring, Body Position, Eye movement, EMG Amplitude, CPAP Pressure, etc
Hypnogram:
Arousals, Respiratory Event, Cardiac Event, Desaturation/SpO2 event, Snoring event,
Limb movements event, etc

Specification

Amplifier :  
Number of Channel : 47 channels
Sampling Rate  : 2000 Hz
Voltage Range : 1 - 12000 μV
Ratio error Voltage Measurement :  
in the range from 10 up to 50 μV  : within ±25%
in the range from 50 up to 10000 μV  : within ±7%
Sensitivity : 0.01 - 10.000.000 μV/mm
Ratio error of sensitivity  : within ±5%
High pass filter : arbitary value can be set in 0.01 - 10 Hz range
Low pass filter  : arbitary value can be set in 0.1 - 500 Hz range
Sweep Speed  : arbitary value can be set in 1 - 10000 mm/s range
Ratio error sweep speed  : within ±2%
Bandpass flatness in 0.5 - 60 Hz  : from -10 up to +5%
Suppresion ratio of power frequency by notch filter : not less than 40 dB
CMRR : not less than 120 dB
Input noise level in 0.5 - 200 Hz range (rms)  : not more than 1.4 μV
Input Impedance : not less than 400 MΩ
Patient leakage current  : not more than 50 nA
Voltage measurement range : 0,2 - 100 mV
General Specifications and Parameters    
Interface  : USB
Supply voltage :  
electronic unit : 5 V DC
desktop PC based system  : 220/230 V AC (50 Hz)
Electronic unit power consumption  : not more than 2.8 VA
Safety  : CF type
Synchronous Review PSG & Video : Include Camera Full HD
Minimum Personal Computer    
i5, 6GB RAM, 500 GB HDD, Windows 10, Speaker Active    
Color Printer & 1 Unit LED Monitor 24"    
Operating Limitations    
Ambient temperature  : 10 to 35 С
Relative humidity  : 80 % at 25 C
Atmospheric pressure  : 760±30 mmHg
Software    
Export format avi, txt, rtf, xml, bmp, emf, EDF, pdf    
Report automatically generated EEG report    
PSG Software    
Recording & Review    
Manual & Automatic Analysis PSG    

Standard Configurations

Neuron-Spectrum-65 electronic unit : 1
LED Photic Stimulator with Stand : 1
Head Caps : 1
Cup EEG electrode with cable : 30
Cable for disposable electrode connection with button clip touch proof : 8
Disposable ECG electrode : 30
Airflow sensor : 1
Snoring Sensor : 4
Inductive thorax and abdominal effort sensor : 1
Body position sensor : 1
Finger sensor for SpO2 recording : 1
Neuron-Spectrum-PU/PSG patient unit : 1
Electrode adhesive paste  : 3
Abrasive paste for skin preparation  : 3
Technical Manual "Neuron-Spectrum PU/PSG" : 1
License for the use of software "Neuron-Spectrum.NET" with additional software module "Neuron-Spectrum.NET/EEG" : 1
License for the use of software "Neuron-Spectrum.NET" with additional software module "Neuron-Spectrum.NET/Video" : 1
License for Neuron-Spectrum.NET software with additional Neuron-Spectrum.NET/PSG software module : 1
     
Aksesoris Lokal:    
PC  : 1
Printer  : 1
Trolley  : 1
UPS  : 1
Camera  : 1