Neuro IOM - 32 Channel with EMG & EP

Description

Neuro-IOM adalah solusi platform modular multifungsi untuk pemantauan neurofisiologis intraoperatif (IONM). Penerapannya memungkinkan tim bedah untuk mendapatkan informasi tentang kemungkinan kerusakan jaringan saraf akibat tindakan ahli bedah dan, dengan demikian, untuk mencegah defisit neurologis. Di sebagian besar negara, pemantauan neurofisiologis intraoperatif adalah suatu keharusan untuk berbagai jenis pembedahan, misalnya, pembedahan tulang belakang, pembedahan otak, pembedahan batang otak, dll. Sedangkan untuk pembedahan tulang belakang (misalnya, skoliosis), taktiknya ditentukan secara langsung selama IONM.

Exellence

32 Amplifier Channels 
8 Channels for electrical stimulation
2 Channels for Low current stimulation
4 Channels for transcranial electrical stimulation
Auditory & Visual stimulator channels
2 ES detector channels
2 Dedicated channels for routine EMG, NCS and EP 

Specification

Number of Channel : 32 channels
Voltage range : 10 µV    150 mV
Ratio error of voltage measurement : ±10%
Common-mode rejection ratio at 50 Hz frequency : > 90 dB
Noise level (with input shorted) : < 0.6 µV in 5 – 3000 Hz
Bandwidth : 0.2 – 4000 Hz @ -3(±0.5) dB
Input impedance (common-mode) : > 300 MΩ
Input capacitance (common-mode) : < 30 pF
Input impedance (differential mode) : > 200 MΩ
DC offset rejection (±800 ± 20 mV) : yes
Patient leakage current : < 0.1 µA
Bandpass flatness in the range:
0.2 to 4000 Hz relative to input signal
: –30 to +5%
Selectable filters @ -3(±10%) dB : yes
High pass filter : 0.2, 0.5, 1, 2, 5, 10, 20, 50, 100, 200, 500, 1000, 2000 Hz
Low pass filter : 10, 15, 20, 30, 35, 50, 75, 100, 150, 200, 300, 500 Hz, 1, 1.5, 2, 3, 4 Kz
Display gain : 0.05, 0.075, 0.1, 0.15, 0.2, 0.25, 0.4, 0.5, 0.75, 1, 1.5, 2, 2.5, 4, 5, 7.5, 10, 15, 20
25, 40, 50, 75,100, 150, 200, 250, 400, 500, 750 µV/div;
1, 1.5, 2, 2.5, 4, 5, 7.5, 10, 15, 20 mV/div
Ratio error of display gain : ±5%
Sweep speed : 0.1, 0.15, 0.2, 0.25, 0.4, 0.5, 0.75, 1, 1.5, 2, 2.5, 4, 5, 7.5, 10, 15, 20, 25, 40, 50, 75, 100, 150, 200, 250, 400, 500, 750 ms/div.;
1, 1.5, 2, 2.5, 4, 5, 7.5, 10, 15, 20 s/div.
Ratio error of sweep speed : ±1%
Sound notification at EMG signal acquisition : yes
Parameters of Electrical StimulatorBuilt-in Electronic Unit    
Number of channels for 32 configurations : 8
Stimulus type : voltage or current
Pulse amplitude at current stimulation : 0 - 200 mA, 0.1 mA step
Maximum voltage at current stimulation : 425 V ± 25 V at 1 kΩ
Pulse amplitude at voltage stimulation : 0 – 400 V, 1 V step
Maximum current at voltage stimulation : 210 mA ± 10 mA  at 1 kΩ
Relative deviation of pulse amplitude : ±5%
Stimulus waveform : rectangular (monophasic, biphasic)
Stimulus polarity : normal, inverse
Pulse duration : 50 - 5000 µs, 10 µs step
Admissible relative deviation of pulse duration in the range : not more than 10%
Stimulation modes : single pulse, train
Stimulation frequency : 0.01 - 100 Hz with 0.05 Hz step
Relative error of stimulation frequency : ± 5%
Number of pulses in train : 2 – 200
Interpulse interval in train : 2 to 10 ms (± 5%) with 0.08 ms step
Admissible absolute deviation of intertrain interval at repetitive stimulation : ± 0.01 ms
Low Current Stimulator Built-in Electronic Unit    
Number of channels for 32 configurations : 2
Current pulse waveform : Rectangular, < 10µ nse time
Pulse polarity : normal, inverse
Pulse amplitude at current stimulation : 0 - 20 mA, 0.01 mA step
Trancranial Electrical Stimulator Built-in Electronic Unit    
Number of Channel 32 Configuration : 1
Number of channel/switched outputs when using electronic switch : 1./4.
Pulse Amplitude : 0 - 1000V, 5 V step
Pulse waveform : Rectangular (Monophasil, Biphasic)
Pulse polarity : Normal, inverse
Stimulation mode : Single Pulse, train, double train, train + pulse
Pulse duration : 40, 50, 60, 70, 80, 100, 150, 200 µs
Auditory Stimulator    
Number of channels 32 Configuration : 2
Parameters : Values
Pulse frequency : 0.01 – 100 Hz
Relative deviation of pulse frequency : ±0.1%
Click duration : 100 – 5000 µs (±20 µs)
Tone frequency : 100 – 5000 Hz
Tone pulse duration : 0.1 – 90 ms
Visual Stimulator    
Number of channels 32 Configuration : 2
Stimulus duration : 2 – 1500 ms
Relative deviation of stimulus duration : ±10%
Stimulus frequency : 0.01 – 100 Hz
Relative deviation of stimulus frequency : ±10%
Maximum luminance of LED goggles per one side : from 2 to 1500 cd/m2
Electrosurgery Detector    
Number of channels 32 Configuration : 2
Minimal detected current within 150 kHz - 2.5 MHz frequency range : 100 mA
Trig in/out Channel    
Number of channels : 1
Input signal : TTL (high – Uin. > 2.4 V; low – Uin. < 0.8 V)
Input impedance : > 47 kΩ
Output signal : TTL (high – Uout. > 4.5 V; low – Uout. < 0.5 V)
Pulse duration : (1 ± 0.01) ms
Output impedance : (1±0.1) kΩ
General Parameters and Specifications    
Interface : USB
Supply voltage : 220±22 V AC (50/60 Hz)
Minimum Computer    
i5, 6GB RAM, 500GB HDD Windows 10 or Higher, 24" monitor & Color Printer    

Standard Configurations

Neuro-IOM 32-channel electronic unit : 1
Neuro-IOM electrical stimulator pod (for 1-4 channels) : 1
Neuro-IOM electrical stimulator pod (for 5-8 channels) : 1
Neuro-IOM low current stimulator pod : 1
Neuro-TES electronic switch : 1
Neuro-IOM acquisition pod (А, 1-8) : 1
Neuro-IOM acquisition pod (B, 9-16) : 1
Neuro-IOM acquisition pod (С, 17-24) : 1
Neuro-IOM acquisition pod (D, 24-32) : 1
Starter Kit IOM  : 1
Starter Kit EMG     
LED photic stimulator : 1
Skate adjustable electro stimulating probe : 1
insert audiometric earphones  : 1
ES detector clip : 1
Video cable  : 1
Video capture device : 1
Medical tape Transpore  : 4
User manual "Neuro-IOM.NET" Soft Copy  : 1
License for IOM : 1
License for EMG  : 1
     
Aksesoris Lokal:    
PC  : 1
Monitor : 2
Laptop    1
Camera  : 1
Printer  : 1
Trolley  : 1
UPS  : 1